Input: 2/3/4-wire PT100, tungsten-rhenium/B/S/N/K/J/T/E thermocouple, 4~20mA, 0~10V, 2~10V, 0~5V, 1~5V.

Output: 6-digit LED display, 4~20mA current range can be set arbitrarily, 0~10V voltage range can be set arbitrarily, relay switch quantity, RS485

Data update rate: thermocouple -1.2s PT100 -1s 4~20mA/0~10V-0.5s

Thermocouple cold end compensation accuracy (AT25℃): (maximum) ±0.05℃

Operating environment: -10~60℃ 0~90% relative humidity without condensation

Special function: Can be matched with the sensor to achieve ultra-high precision measurement. When thermocouple input, four-wire PT100 can be connected externally

Power supply: 220VAC/24VDC

Product Specificatiopns

Accuracy (AT25℃ Max):

Enter model number

Range (℃)

Measurement accuracy

Impact of environment change of 1 ℃ (temperature drift)

4-wire system PT100

-200 ~ 500

±0.02 ℃

0.003 ℃

3-wire system PT100

-200 ~ 500

±0.05 ℃

0.003 ℃

2-wire system PT100

-200 ~ 500

±0.05 ℃

0.003 ℃

B-Type thermocouple

300 ~ 1820

±1.20 ℃

0.036 ℃

S-type thermocouple

0 ~ 1768

±0.60 ℃

0.029 ℃

N-type thermocouple

-200 ~ 1300

±0.32 ℃

0.025 ℃

K-type thermocouple

-180 ~ 1372

±0.22 ℃

0.024 ℃

J-type thermocouple

-180 ~ 760

±0.18 ℃

0.008 ℃

T-Type thermocouple

-200 ~ 400

±0.18 ℃

0.011 ℃

E-type thermocouple

-50 ~ 1000

±0.15 ℃

0.007 ℃

Note 1: Total digital accuracy of thermocouple measurement: analog/digital conversion accuracy +0.05℃ (cold end compensation accuracy)

Overall accuracy after using sensor matching function (transmitter + sensor)

Enter model number

Range (℃)

A/D conversion accuracy (℃)

PT100

-50 ~ 200

± 0.02

S-type thermocouple

0 ~ 650

± 0.70

650 ~ 1100

± 1.50

N-type thermocouple

0 ~ 650

± 0.40

650 ~ 1100

± 1.90

T-Type thermocouple

-200 ~ -70

± 0.40

-70 ~ 400

± 0.20

Note 1: The sensors listed in the table are the most commonly used, stable and representative sensors in each temperature range. For details of accuracy in a wider temperature range, please consult the manufacturer.

Note 2: The thermocouple matching function is a patented technology and must be completed before leaving the factory.

Note 3: The accuracy of thermocouples in the table includes the accuracy of cold junction compensation and does not need to be superimposed separately.