WDet-7000TPN Online Nitrate Nitrogen Water Quality Analyzer

Negotiable
China
Room 301, Floor 3, Building 1, No. 22 Zhiren Street, Puyan Subdistrict, Binjiang District, Hangzhou, Zhejiang Province
0086-571-28322780
Online Contact
Favorite

Product Introduction

Product Overview

The WDet-7000TPN Online Nitrate Nitrogen Water Quality Analyzer is a newly developed online nitrate nitrogen monitoring instrument based on our company's patented new metering technology platform, combined with self-developed integrated spectral measurement technology. The working principle of this product complies with "HJ/T346—2007 Water Quality—Determination of Nitrate Nitrogen—Ultraviolet Spectrophotometry," featuring high reliability, low limit of quantification, and excellent stability.

Product Features

  • Adopts a novel quantification technology resistant to color and turbidity interference; integrated valve island reduces dead volume, easy to disassemble and clean, with low-cost maintenance.
  • Built-in multi-range adaptive adjustment; fully automatic calibration and cleaning without manual intervention.
  • Equipped with self-diagnosis and safety protection functions; real-time leak alarm and lockout; complete fault information recording for easy query.
  • Supports remote monitoring, upgrade, and maintenance via ETHERNET/GPRS and other interfaces.
  • Features a 7-inch Android touchscreen with intuitive interface and smooth operation.
  • Provides multiple industrial communication and control interfaces, with flexible function expansion as needed.

Application Fields

Suitable for pollution sources, industrial process water, municipal wastewater, and other fields.

Product Parameters

Test Method Ultraviolet Spectrophotometry Measurement Range 0~5/30/100mg/L or 0~5/30/200mg/L
Low Concentration Drift ≤0.05mg/L High Concentration Drift ≤1%
Repeatability ≤2% Indication Error 20%FS, ±10%; 50%FS, ±8%; 80%FS, ±5% (0~50mg/L)
Limit of Quantification ≤0.3mg/L Measurement Interval On-the-hour measurement, interval measurement, also can trigger the instrument via Modbus
Voltage Influence Test ±5% Ambient Temperature (5~40)℃, (65±20)%RH, non-condensing
Minimum Maintenance Cycle ≥168h/time Power Supply AC(220±22)V, (50±0.5)Hz
Calibration Supports manual and automatic calibration. Calibration cycle and time are configurable Equipment Power 200W
Communication Interface Output: 1 channel (4-20)mA, RS232/RS485, 2 multifunctional relays Dimensions 412.5mm*323.5mm*562mm (L*W*H)
Input: 1 channel (4-20)mA, 2 digital inputs Weight 25kg