JSW-Series Walk-In Fuel Cell Stack Environmental Test Chamber

Negotiable
China
18030810556
Online Contact
Favorite

Product Introduction

Product Features

◆ Provides precise climate environment testing solutions for hydrogen fuel cell stacks of different power ratings.

◆ Highly targeted products covering the vast majority of medium and large fuel cell stacks on the market.

◆ Integrated design incorporating the required hydrogen/air pre-cooling system and glycol circulation system.

◆ Advanced hardware-software integration ensures close coordination between the environmental test chamber and the test bench.

◆ High safety with a safety matrix interlocking response strategy to ensure safe testing operations.

Model JSW-8T JSW-15T JSW-24T
Compatible Fuel Cell Stack 80KW 120KW 150KW 200KW 250KW 300KW
Hydrogen/Air Pre-Cooling Unit JHAP26-T JHAP39-T JHAP412-T JHAP515-T JHAP618-T
Internal Volume 8m³ 15m³ 24m³
Internal Dimensions (D×W×H) 2000mm*2000mm*2000mm 2500mm*2500mm*2400mm 3200mm*3000mm*2500mm
Chamber External Dimensions (D×W×H) 3300mm*2300mm*2700mm 3800mm*2800mm*3100mm 4500mm*3300mm*3200mm
Refrigeration Unit Dimensions (D×W×H) 1700mm*1350mm*1700mm 2200mm*1200mm*2000mm 2200mm*1400mm*2000mm
Hydrogen/Air Pre-Cooling Dimensions (D×W×H) 2250mm*1250mm*2600mm 2400mm*1400mm*2800mm
Pressure Relief Opening Size 0.24m² 0.36m² 0.45m²
Installed Power 90KW 113KW 155KW 170KW 183KW
Maximum Current 135A 180A 248A 278A 304A
Cooling Water Flow Rate 20m³/h 26m³/h 36m³/h 39m³/h 43m³/h
Temperature Range -40°C~90°C
Heating Rate ≥1°C/min (within -40°C~90°C range, full-process average, with loaded hydrogen fuel cell stack and tooling, under no heat generation).
Cooling Rate ≥1°C/min (within -40°C~90°C range, full-process average, with loaded hydrogen fuel cell stack and tooling, under no heat generation).
Temperature Deviation ≤±2.0°C (no load)
Temperature Fluctuation ≤±0.5°C (static, no thermal load) ≤±2.0°C (dynamic test conditions)
Temperature Uniformity ≤2.0°C (no load, normal pressure, static)
Humidity Range 20% RH~95%RH
Temperature Range During Humidity Testing 20°C~85°C
Humidity Deviation (Humidity ≥75%RH) +2~-3%RH (Humidity <75%RH) ±5%RH