Table 1: Product Selection
|
Version Model |
Air pressure version |
Differential pressure version |
Micro differential pressure version |
|||
|
Functional features |
810 |
810-S |
810 |
810-S |
810 |
810-S |
|
Automatic pressure generation range① |
(-0.085~2.5)MPa |
(-85~250)kPa |
± 10kPa |
|||
|
Optional range of pressure control module② |
Minimum 60kPa Level 0.01/Level 0.02 |
Minimum ± 10kPa Level 0.01/Level 0.02 |
Minimum ± 500Pa Level 0.02/Level 0.05 |
|||
|
Control stability③ (30s) |
<0.005%FS |
<0.005%FS |
<0.005%FS |
|||
|
Target pressure stabilization duration |
> 5min |
> 5min |
> 5min |
|||
|
Control response time④ |
<40s |
<20s |
<30s |
|||
|
Pressure type |
Gauge pressure (switchable to absolute pressure⑤) |
Differential pressure |
Differential pressure |
|||
|
External CDP module to measure pressure |
● |
● |
● |
● |
● |
● |
|
Data storage |
● |
● |
● |
● |
● |
● |
|
Task Management |
● |
● |
||||
|
HART |
● |
|||||
① The actual pressure automatic occurrence range depends on the pressure control module; the negative pressure part occurrence range is an indicator when the ambient atmospheric pressure is 100kPa.a;
② At least one CDPM pressure control module shall be configured, which can be replaced. See the selection table for details;
③ The control stability is based on the range of the pressure control module;
④ When no additional load volume is applied and 20% of the range step is completed;
⑤ Optional function: based on the internal atmospheric pressure module, it can cooperate with CDPM to realize free switching of gauge pressure/absolute pressure. the internal atmospheric pressure module can be calibrated with a measurement range of (60~110)kPa.a,
Accuracy is ± 55Pa.
Table 2: Electrical measurement indicators (ambient temperature 5℃ ~ 35℃,1 year accuracy)
|
Project |
Scope |
Resolution |
Indicators |
|
Voltage measurement |
(-30~30)V |
0.1mV |
±(0.01%RD+1.5mV) |
|
Current measurement |
(-30~30)mA |
0.1μA |
±(0.01%RD+1.5μA) |
|
Switch detection |
Supports mechanical and NPN/PNP digital |
||
|
Current output |
(0~24)mA |
lμA |
±(0.01%RD+1.2μA) |
|
24V output |
Maximum load capacity 50mA, ripple peak ≤ 40mV |
||
|
Note: Exceeding temperature coefficient: +(0.001%RD +0.001%ES)/℃ |
|||
Table 3: Product Parameters
|
Project |
Details |
|
Power supply |
Lithium battery or special adapter. The typical working time of the battery is 10 hours, the charging time is 4 hours, and it can be charged independently. |
|
Current measurement |
The maximum number of continuous storage is 1 million, and the maximum number of tasks is 1000. |
|
Environmental Specifications |
Operating temperature:(0~50)℃; Environmental relative humidity: ≤ 90%, non-condensing; Altitude:<3000m; Storage temperature:(-20~70)℃ |
|
Communication mode |
USB, WiFi, Bluetooth (BLE) |
|
Overall dimension |
Height 238mm × Width 110mm × Thickness 70mm |
|
Weight |
Less than 1.8kg |


京公网安备 11010802043282号