ConST810 Handheld Fully Automatic Pressure Calibrator
Negotiable
China
Building 5, No. 3 Fengxiu Middle Road, Haidian District, Beijing
+86 010-56973333
Online Contact
Favorite
Product Introduction

Table 1: Product Selection

Version Model

Air pressure version

Differential pressure version

Micro differential pressure version

Functional features

810

810-S

810

810-S

810

810-S

Automatic pressure generation range

(-0.085~2.5)MPa

(-85~250)kPa

± 10kPa

Optional range of pressure control module

Minimum 60kPa

Level 0.01/Level 0.02

Minimum ± 10kPa

Level 0.01/Level 0.02

Minimum ± 500Pa

Level 0.02/Level 0.05

Control stability (30s)

<0.005%FS

<0.005%FS

<0.005%FS

Target pressure stabilization duration

> 5min

> 5min

> 5min

Control response time

<40s

<20s

<30s

Pressure type

Gauge pressure (switchable to absolute pressure)

Differential pressure

Differential pressure

External CDP module to measure pressure

Data storage

Task Management

 

     

HART

   

     

① The actual pressure automatic occurrence range depends on the pressure control module; the negative pressure part occurrence range is an indicator when the ambient atmospheric pressure is 100kPa.a;

② At least one CDPM pressure control module shall be configured, which can be replaced. See the selection table for details;

③ The control stability is based on the range of the pressure control module;

④ When no additional load volume is applied and 20% of the range step is completed;

⑤ Optional function: based on the internal atmospheric pressure module, it can cooperate with CDPM to realize free switching of gauge pressure/absolute pressure. the internal atmospheric pressure module can be calibrated with a measurement range of (60~110)kPa.a,

Accuracy is ± 55Pa.

 

Table 2: Electrical measurement indicators (ambient temperature 5℃ ~ 35℃,1 year accuracy)

Project

Scope

Resolution

Indicators

Voltage measurement

(-30~30)V

0.1mV

±(0.01%RD+1.5mV)

Current measurement

(-30~30)mA

0.1μA

±(0.01%RD+1.5μA)

Switch detection

Supports mechanical and NPN/PNP digital

Current output

(0~24)mA

lμA

±(0.01%RD+1.2μA)

24V output

Maximum load capacity 50mA, ripple peak ≤ 40mV

Note: Exceeding temperature coefficient: +(0.001%RD +0.001%ES)/℃

 

Table 3: Product Parameters

Project

Details

Power supply

Lithium battery or special adapter. The typical working time of the battery is 10 hours, the charging time is 4 hours, and it can be charged independently.

Current measurement

The maximum number of continuous storage is 1 million, and the maximum number of tasks is 1000.

Environmental Specifications

Operating temperature:(0~50)℃;

Environmental relative humidity: ≤ 90%, non-condensing;

Altitude:<3000m;

Storage temperature:(-20~70)℃

Communication mode

USB, WiFi, Bluetooth (BLE)

Overall dimension

Height 238mm × Width 110mm × Thickness 70mm

Weight