Product Introduction

Product Features:

Equipped with a 10-inch color capacitive touchscreen across the entire series, offering an enhanced user experience and improved touch operation convenience

Features an independently developed intelligent control logic system for precise control of temperature and humidity curves, with a wide operating range (see Humidity MAP), meeting diverse testing requirements

The inner chamber is made of SUS304 mirror-finish stainless steel, providing excellent corrosion resistance, easy-to-clean surfaces, and simplified maintenance

A unique air circulation system and software control system significantly reduce water consumption under high temperature and humidity conditions (without water recycling); water usage per unit time under equivalent conditions is among the lowest in the domestic market

Advanced energy-saving design with an intelligent control system and specialized air circulation system that automatically adjusts operating parameters for efficient energy utilization

Product Parameters

Model

JC330(C)

JC610(C)

JC1000(C)

JT1500(C)

JT2000(C)

Nominal internal volume (L)

330L

610L

1000L

1500L

2000L

Internal dimensions (mm)

D

760

795

960

1470

1670

W

585

800

1100

1100

1100

H

750

950

950

950

1100

External dimensions (mm)

D

1820

1850

2000

2525

2700

W

880

1100

1450

1450

1430

H

1950

2230

2230

2230

2400

Performance parameters

Minimum temperature

Minimum temperature with C at the end -70 ℃/minimum temperature without C -40 ℃

Maximum temperature

+180 ℃ (the maximum temperature can be up to 220 ℃ according to the actual needs of users)

Temperature fluctuation

± 0.1 ℃ ~ ± 0.5 ℃

Temperature uniformity

≤ 2 ℃(≤ 150 ℃)

Temperature deviation

≤ ± 2 ℃

Heating rate

C/min

4.0

6.0

4.0

4.0

3.0

Cooling rate

℃/min

3.5

2.6

2.8

2.5

2.0

Humidity range

%R.H.

10% ~ + 98% R.H (humidity working range +10 ℃ ~ + 95C, see temperature and humidity MAP chart for details)

Dew point temperature range

0 to 94 ℃

Humidity deviation

%R. H.

-3%r.h.~+2%r.h

Humidity fluctuation

%R. H.

≤±3%R.H.

Test Method

The above indicators are measured at no-load, ambient temperature 20±2 ℃ (according to GB/T5170.2-2017) rate assessment range: Tmin-Tmax, according to IEC60068-3-5, measured at the air outlet, empty container indicators. Temperature and humidity performance test According to the standard IEC60068-3-5, the sensor is placed at the air outlet of the air handling unit.

Meet the standard

Low temperature test: IEC60068-2-1; IEC60721-4; MIL-STD-810G, Meth. 502.5; GB/T2423.1TestA

High Temperature Test: IEC60068-2-2; IEC60721-4; MIL-STD-202G, Meth. 108A; MIL-STD-810G, Meth. 501.5; MIL-STD-883J, Meth. 1008.2; GB/T2423.2TestB

Temperature and humidity test: IEC60068-2-67; IEC60068-2-78; IEC60721-4; ISO16750-4, Dampheatsteady; GB/T2423.3TestCab; GB/T2423.4TestDb-Testcycle-Variant1 and Variant2; GB/T2423.50TestCy; IEC60068-2-30TestDb-Testcycle-Variant1 and Variant2

Thermal compensation capability

1500~3000W 400W (temperature 25 ℃ to 90 ℃, humidity not exceeding 90%)

Cooling mode

Air/water cooling (optional)

Power supply form

Voltage: 380V ± 10% three-phase five-wire system; frequency: 50Hz ± 10%

Maximum power

KW

4.8/6.6

8.5/10.5

13.5/16

15.5/18

20/23

Maximum current

A

13/15

16/21

23.5/29

24/30

39/45

Noise level

Db(A)

65

68

68

68

70

Equipment weight

KG

580/625

680/750

950/1050

1200/1350

1400/1500