Model
|
JTS50M-B/S/E
|
JTS100M-B/S/E
|
JTS150M-B/S/E
|
JTS200M-B/S/E
|
JTS320M-B/S/E
|
JTS500M-B/S/E
|
JTS1000M-B/S/E
|
|
Inner volume of basket (liter)
|
50L
|
100L
|
150L
|
200L
|
320L
|
500L
|
1000L
|
|
Performance parameters
|
Regenerator
|
Temperature exposure range
|
+50 ℃ ~ +150 ℃
|
|
Upper preheat temperature limit
|
200 ℃
|
|
Heating time
|
RT ~ +200 ℃, B type ≤ 50min/S type ≤ 40min/E type ≤ 30min
|
|
Cold storage room
|
Temperature exposure range
|
Type B 0 ~-40 ℃/Type S 0 ~-55 ℃/Type E 0 ~-65 ℃
|
|
Lower precool temperature limit
|
-75 ℃
|
|
Cooling time
|
RT ~-70 ℃: B type ≤ 60min/S type ≤ 50min/E type ≤ 40min
|
|
Working room
|
Temperature shock range
|
Type B -40 ℃ ~ +150 ℃/Type S -55 ℃ ~ +150 ℃/Type E -65 ℃ ~ +150 ℃
|
|
Temperature fluctuation
|
≤ ± 0.5
|
|
Temperature uniformity
|
≤ 2 ℃
|
|
Temperature deviation
|
≤ ± 2 ℃
|
|
Door switching time
|
≤ 10s
|
|
Temperature recovery time
|
≤ 5min
|
|
Load (steel)
|
3kg
|
5kg
|
7kg
|
10kg
|
15kg
|
20kg
|
40kg
|
|
Optional: high and low temperature test function
|
Temperature range
|
-65 ℃ ~ +150 ℃
|
|
Temperature rise and fall rate
|
Type B is about 2 ℃/min, Type S is about 3 ℃/min, Type E is about 4 ℃/min
|
|
Remarks
|
The above indicators are measured under no-load and ambient temperature of 20±2 ℃ (according to GB/T 5170.2-2017)
|
|
Working room size (mm)
|
D
|
350
|
400
|
500
|
600
|
650
|
800
|
|
|
W
|
350
|
500
|
600
|
600
|
700
|
700
|
800
|
|
H
|
400
|
500
|
500
|
550
|
700
|
900
|
1000
|
|
External dimensions (mm)
|
D
|
2000
|
2250
|
2350
|
2550
|
1950
|
2200
|
2650
|
|
W
|
1400
|
1570
|
1670
|
1670
|
2700
|
3000
|
3000
|
|
H
|
1950
|
2150
|
2150
|
2250
|
2300
|
2400
|
2600
|
|
Caster
|
4, more than 500L for 6 (the above dimensions do not include casters, casters increased by 60-120mm)
|
|
Test hole
|
1 φ50
|
1 φ100
|
|
Stainless steel shelf (2 standard)
|
Bearing 15kg/block bearing 30kg block bearing 50kg/block
|
|
Use ambient temperature
|
0 ~ +40 ℃
|
|
Shell
|
Plastic-sprayed steel plate
|
|
Inner wall
|
SUS304 stainless steel plate
|
|
Cooling mode
|
Water cooling (water temperature 7~28 ℃, water pressure 0.2~0.4Mpa, in order to ensure the cooling performance)
|
|
Controller
|
7' color touch screen controller
|
|
Communication function
|
RS485 or standard RJ45 network interface, standard USB interface
|
|
Power
|
80V 3±10%/50Hz ± 1, three-phase five-wire system (3P + N + G)
|
|
Installed Capacity (kw)(B/S/E)
|
14/17/22
|
17/22/29
|
23/28/34
|
27/33/40
|
33/40/50
|
50/60/75
|
50/60/75
|
|
Box weight (kg)
|
720
|
1200
|
1500
|
1700
|
2200
|
2500
|
3200
|
|
Meet the test method
|
GJB150.52009GB/T24232008; After the high and low temperature test function is selected, it also meets the requirements of the high and low temperature test.
|
|
Safety protection measures
|
Power supply undervoltage, overvoltage, phase loss, phase sequence error protection; Triple overtemperature protection; Heater short circuit protection, refrigerator overpressure/overload/oil shortage protection; Humidifier fault protection; Blower motor overload/overheat protection; Control board overcurrent protection, leakage protection; Cooling water shortage protection, gas source undervoltage protection, etc.
|
|
Remarks
|
1. Optional high and low temperature test function; 2. Optional special test holes and special shelves; 3. Optional liquid nitrogen function
|