Lao Ying Model 5032 Fully Automatic Permanganate Index Analyzer

Negotiable
China
No.379, Hedong Road, High-tech District, Qingdao, Shandong, China 266109
13706305199
+86-18221933371
Online Contact
Favorite

Product Introduction

 
   

This instrument is specifically designed for the determination of the permanganate index (CODMn value) in water quality samples. Its principle is based on the oxidation of reducing substances by potassium permanganate in acidic or alkaline solution, with the excess potassium permanganate titrated and reduced by sodium oxalate solution, and the permanganate index expressed in terms of potassium permanganate consumption.

It can be used by environmental protection, health, labor, safety supervision, military, scientific research, education and other departments for the determination of the permanganate index.

 

 Standards Implemented

  • GB11892-89 Water quality — Determination of permanganate index
  • GB/T 5750.7-2023 Standard examination methods for drinking water — Part 7: Comprehensive indicators of organic matter

 

 Main Features

  • Adopts fully automatic integrated technology, enabling fully automatic digestion of water samples, one-key detection, unattended operation, and direct output of titration results.
  • Adopts an intelligent robotic arm for precise and stable transfer of sample cups, replacing manual labor and enabling high-temperature cup gripping and high-temperature titration, with "digestion-titration" completed fully automatically.
  • Adopts a high-precision visual sensor that simulates the human eye's visual recognition system, automatically determining the titration endpoint based on color changes, with accurate and reliable endpoint discrimination results.
  • Each type of reagent is metered independently, with reagent consumption monitoring and low-level warning alarms to alert the user, ensuring effective operation of the experiment.
  • The data workstation features a schematic diagram matching the sample positions; samples can be freely withdrawn, added, or replaced during operation without stopping or pausing the system, making operation more user-friendly.

Product Parameters

Main Parameters

Parameter Range

Sample tray

54 positions

Heating method

Water bath heating, 9 digestion positions, intelligent staged temperature-controlled heating

Titration channels

Dual channel

Titration determination

RGB color recognition

Precision

RSD≤3.0% (glucose standard solution at 3.0 mg/L, n=3)

Measurement range

(0~6.0) mg/L (without dilution, at 100 mL sample volume)

Measurement speed

≤4 minutes/sample

Temperature control accuracy

±1℃

Reagent safety monitoring

Four-reagent monitoring, real-time display of reagent levels

Drive robotic arm

Electric control