JT-11/JT-11A Ultra-precision Handheld Digital Thermometer
Negotiable
China
No. 212, Jinyi Street, Taihe District, Jinzhou City, Liaoning Province
+86 0416-2327769
Online Contact
Favorite
Product Introduction

Input type: PT20/PT100/PT200/PT500/user-defined thermal resistor B/R/S/N/K/J/T/E thermocouple/thermopile

Number of input channels: 2

Main functions: sensor matching to achieve ultra-high precision measurement.

Maximum and minimum values, average value, maximum-minimum value display.

Over-limit alarm.

℃/F/K/Q/mv measurement unit switching.

Data storage up to 390,000 records, including time information.

USB data directly connected to computer.

Power supply: 6~15VDC external power supply or 4 AAA batteries.

Communication rate: 9600bps (JT-11)

1200/2400/4800/9600/19200/38400bps can be set (JT-11A)

Purpose:

①Precision resistance detection, platinum resistance temperature corresponding resistance detection saves the consumption of second-class standard thermometers.

②Constant temperature field acceptance.

Product Specifications

Resistance measurement accuracy (AT25 ℃):

Range

Sampling mode

Resolution (typical) Ω

Error (Max) Ω

300Ω

Ω

0.0001

±(0.0005+0.0006%*READ)

TRUEQ

3kQ

Ω

0.006

±(0.01+0.005%*READ)

TRUEQ

DC voltage measurement accuracy (AT25 ℃)

Range mV

Resolution (typical) μV

Error (Max) μV

70

0.1

±(1+0.004%*READ)

Resistance measurement accuracy (AT25 ℃): Temperature Drift

Project

Impact per 1 ℃ change of environment (maximum)

Resistance

2ppm*READ

Voltage

7ppm*READ

Resistance measurement accuracy (AT25 ℃): electrical performance

The parameter conditions given in this table are: ambient temperature 25 ℃, power supply voltage 6V, dual-channel resistor sampling, and measuring range 300 Ω.

Project

Condition

Typical value

Working current

Turn on backlight

13.5mA

Turn off backlight

11mA

Resistance measurement sampling time

Step change> 3mΩ, two channels simultaneously measure resistance

High resolution

Low resolution

6s

2s

Step change <3mΩ, two-channel simultaneous measurement of resistance

High resolution

Low resolution

18s

6s

Voltage measurement sampling time

Two-channel simultaneous measurement of mv

0.5s

Thermocouple measurement sampling time

Simultaneous measurement of thermocouple signals with two channels

High resolution

Low resolution

6.5s

2.5s

Resistance measurement accuracy (AT25 ℃): the overall accuracy after matching with the sensor

Enter model number

Range (℃)

Accuracy (℃)

PT100

-50 ~ 200

± 0.01

PT50

-60 ~ 400

± 0.02

T-Type thermocouple

-200 ~ -50

± 0.3

-50 ~ -300

± 0.15

N-type thermocouple

0 ~ 400

± 0.3

300 ~ 600

± 0.5

S-type thermocouple

0 ~ 600

± 0.5

600 ~ 1100

± 1.5