924 Toxic Air Sampler
Negotiable
China
**No. 611, Building 6, Bonded Logistics Park, No.777 Guanggu 3rd Road, Donghu New Technology Development Zone, Wuhan
027-87926078
Online Contact
Favorite
Product Introduction

Product Overview:

The 924 Toxic Air Sampler is used for unattended collection of ambient air, followed by laboratory analysis of toxic components. The sampler adopts a modular design, consisting of a control unit, pump box assembly, weatherproof enclosure, and sampling head device. The sampler can be equipped with 1-4 sampling channels, including 37/47mm filters (for collecting toxic metals and hexavalent chromium in particulate matter), adsorption tubes (for collecting carbonyl compounds in air), and sampling canisters (for collecting VOCs in air). In addition to the sampling canister channel, the sampler can be equipped with multiple sampling channels. The sampler can only connect to one sampling canister.

Product Features:

  • Multiple configurations including sampling canisters, carbonyl compounds, 37/47mm filters, and adsorption tubes
  • Weatherproof independent 4-channel sampler
  • Corrosion-resistant stainless steel and brass fittings
  • Integrated computer, printer, and touchscreen display
  • Equipped with heater and fan, suitable for outdoor weather operation (temperature: 5-50℃, relative humidity: 0-90%, non-condensing)
  • Modular design, convenient for assembly, installation, operation, and maintenance
Product Parameters

● Sampling flow control

Mass flow controller, typical 0-2L/min, 0-30L/min or customized, accuracy and linearity up to ± 1% full scale

● Display

Touch screen, TFT color graphic display, white LED backlight, diagonal 5.7 ", pixel 320 x 240

● External sampling pump

Gast1023 series motor (vane type)

 

Maximum vacuum 26 "Hg

 

Unobstructed flow 283L/min

● Digital vacuum sensor

30 "Hg - 0psig, accuracy ± 0.5% full scale (best fit straight line)

● Integrated temperature control

Built-in temperature control circuit, through 100W silicone rubber heater control chassis temperature

● Temperature display

Display chassis temperature, Honeywell resistance sensor, -40°C - 150°C

● Power

115V/60Hz; 220V/50Hz (optional)

● Dimensions

7'(H)×1'10”(W) ×1'(D); 2134mm(H)x559mm(W)x305mm(D)

● Weight

 

Sampler (excluding pump)

Approximately 58kg (128lbs)

Sampling pump

Approximately 23.6kg (52lbs)