X-Orifice Orifice Flow-limiting Constant Flow Sampler
Negotiable
China
**No. 611, Building 6, Bonded Logistics Park, No.777 Guanggu 3rd Road, Donghu New Technology Development Zone, Wuhan
027-87926078
Online Contact
Favorite
Product Introduction

Product Overview:

The flow-limiting orifice sampler is an instrument that uses an orifice to limit gas flow and achieves constant flow sampling through a flow controller. It is mainly used for unattended on-site VOCs sampling, and can also be used for sampling ambient air and other pollution source environmental gases.

Product Applications:

  • Ambient air sampling
  • Indoor air analysis
  • Pollution source perimeter monitoring
  • U.S. EPA methods TO-12, TO-14A, TO-15 compound analysis
  • Select corresponding flow range and orifice based on SUMMA canister volume and sampling duration

Product Features:

  • Flow controller ensures relatively constant flow
  • Filter to remove particulate matter from ambient air
  • Vacuum pressure gauge to test SUMMA canister pressure before and after sampling
  • Different orifices configured according to various sampling flow requirements
Product Parameters

Restriction orifice diameter

In. (inch)

Restriction orifice flow rate range

cc/min

SUMMA tank volume/corresponding sampling time

1L

3L

6L

15L

0.0008

0.5-2

24 hours

48 hours

125 hours

/

0.0012

2-4

4 hours

12 hours

24 hours

60 hours

0.0016

4-8

2 hours

6 hours

12 hours

30 hours

0.0020

8-15

1 hour

4 hours

8 hours

20 hours

0.0030

15-30

/

2 hours

3 hours

8 hours

0.0060

30-80

/

/

1.5 hours

4 hours

0.0090

80-340

/

/

0.5 hours

1 hour