TH-2008H Multi-Gas Dynamic Calibrator

EXW
$0.15-0.29 /Unit
China
No.11 Huashiyuan North Road, East Lake High-tech Development Zone, Wuhan City, Hubei Province
02787923722
Online Contact
Favorite
Product Introduction

Working Principle

TH-2008H Multi-Gas Dynamic Calibrator is used to calibrate various ambient air analyzers. It has 2 built-in mass flow controllers (standard configuration) to control zero gas and standard gas flow rates respectively. Used in conjunction with a zero air generator and various standard gases, it calibrates SO2, NO, CO analyzers using the method of diluting standard gas with zero air. It has a built-in ozone generator and detection unit to generate O3 gas. The O3 concentration is calibrated in 16 segments and can generate standard ozone gas at fixed or arbitrary concentrations to calibrate ozone analyzers. It can generate quantitative NO2 standard gas and use gas phase titration to calibrate the NO2→NO conversion furnace efficiency in NOx analyzers. Through preset programming or user random operation, it completes calibration of analyzers. After all calibrations are preset, one key completes output of various standard gases requiring quantitative output such as SO2, NO, NO2, O3, CO, etc.

Instrument Features

•   Adopts 7-inch full-touch color screen;

•   Chinese menu-style operation interface, easy to operate;

•   With nearly a hundred program segments and sequence segments that can be set to flexibly preset various parameters for instrument calibration;

•   With automatic temperature and pressure compensation function;

•   Can accept various remote zero and span calibration commands, and remotely diagnose internal working parameters;

•   Can query settings of program segments and sequence segments;

•   With multi-parameter alarm function;

•   With RS232, RS485, USB and Ethernet data transmission functions.

Product Parameters

Dilution Gas Inlet Ports

1

Input Gas Pressure

(100~200)kPa

Standard Gas Inlet Ports

3

Response Time (T95)

≤60S

Dilution Gas Mass Flow Controller

(0~10)L/min (expandable)

Dilution Ratio (Standard)

1:100~1:1000

Standard Gas Mass Flow Controller 1

(0~100)mL/min (expandable)

Output Channels

3

Standard Gas Mass Flow Controller 2

(0~20)mL/min (optional)

Ozone Generation Range

(25~1000)ppb

Flow Linearity Error

±1%

Ozone Repeatability

≤1%

Flow Repeatability

≤0.5%

Operating Ambient Temperature

(15~35)℃

Dimensions

(482×570×222) mm

Weight

23kg