Instrument Principle
The instrument uses dedicated chromatographic column combination, center cutting with backflush technology, and hydrogen flame ionization detector (FID) technology to detect methane and non-methane total hydrocarbons. After the sample is collected into the quantitative loop, the injection valve is switched, and under carrier gas drive, it enters the chromatographic columns for separation. One path of total hydrocarbons enters the FID detector for detection, the other path of oxygen and methane is separated from the PQ chromatographic column, enters the 5A chromatographic column, after switching valve position, hydrocarbons are backflushed, methane enters the FID detector for detection.
Instrument Features
•Uses center cutting with backflush technology, complies with standard HJ/T38-1999, directly measures methane and non-methane total hydrocarbons;
•Uses dedicated chromatographic column combination, sample insulated at 180℃, peaks within 30s, no residue, high sensitivity;
•Instrument has power-on self-test function, gas interruption protection function, short-term power failure restart function and alarm function, ensuring system safety and stability;
•FID detector has automatic ignition function and wide range output, linear range 10-7;
•Uses automatic electronic flow control technology (EPC) to control carrier gas, air and hydrogen, high precision (0.01psi), good repeatability and reproducibility;
•Core components use internationally renowned brands, high reliability, long service life.
Application Fields
Ambient air online monitoring or scientific research, petrochemical, spraying, printing and other emission outlets or plant boundary online monitoring.