TH-150H Medium Flow Ambient Particulate Matter Sampler

EXW
$0.14-0.29 /Unit
China
No.11 Huashiyuan North Road, East Lake High-tech Development Zone, Wuhan City, Hubei Province
02787923722
Online Contact
Favorite
Product Introduction

The product complies with GB/T 15432-1995 Gravimetric Determination of Total Suspended Particulate Matter in Ambient Air; JJG 943-2011 Verification Regulation for Total Suspended Particulate Matter Samplers; HJ/T374-2007 Technical Requirements and Detection Methods for Total Suspended Particulate Matter Samplers; HJ/T 375-2007 Technical Requirements and Detection Methods for Ambient Air Samplers; Enterprise standard Q/WTH 04Y-2023 TH-150 Series Medium Flow Ambient Particulate Matter Samplers.

Application Scope

The TH-150H medium flow ambient particulate matter sampler is a new medium flow particulate/gas combination sampler developed by our company with independent intellectual property rights. It uses a new energy-saving brushless pump and built-in battery to maintain 100L/min flow for continuous operation. It can complete sampling without stopping during mains power failure, making it an ideal instrument for environmental assessment and on-site monitoring.

Instrument Features

•Three-channel sampling flow with automatic constant flow and digital flow display, avoiding disadvantages of rotor flowmeters for gas sampling such as manual flow adjustment, inaccurate flow, and float drop;

•Uses original imported built-in suction pump with high reliability and strong load resistance;

•Configured with imported flow sensor to detect sampling flow, with microcontroller automatically compensating for flow changes caused by voltage and resistance variations for more accurate flow measurement;

•The instrument can sample one channel of particulate matter and two channels of gaseous pollutants simultaneously or separately;

•4.3-inch color touch screen for display and operation, clearly visible in sunlight;

•Three channels independently accumulate working condition sampling volume and reference state volume, with separate setting of sampling time and sampling flow for each channel;

•Features software calibration function, allowing convenient calibration of parameters such as instrument flow, temperature, and atmospheric pressure via touch screen;

•The instrument has storage, query, and printing functions (printer optional), with each channel capable of storing more than 200 groups independently;

•The instrument is small in size, light in weight, and easy to carry;

•Configured with DC battery, can operate for more than 20 hours (particulate sampling only, domestic glass fiber filter, flow 100L/min), meeting 24-hour valid sample requirements;

•External power interface.

Product Parameters

Item

Particulate Sampling

Gas Sampling

Flow Range

(60~130) L/min

(0.1~2.0) L/min

Resolution

0.1 L/min

0.001 L/min

Indication Error

±2%

±2%

Stability

±5%

±5%

Repeatability

≤2%

≤2%

Atmospheric Pressure Measurement Range

(60~110) kPa

Resolution

0.01 kPa

Indication Error

±0.5 kPa

Temperature Measurement Range

(-20~50) ℃

Indication Error

±2.0 ℃

Timing Error

±0.1%

Charging Power Supply

AC220(1±10)% V (50±1)Hz

Main Unit Weight

6kg (including built-in battery)

Main Unit Dimensions

(160×300×260)mm (L×W×H)

Cutter Technical Specifications

PM₂.₅ Cutter

Cut Size

Da₅₀=(2.5±0.2)μm

 

Collection Efficiency

σg=1.2±0.1

PM₁₀ Cutter

Cut Size

Da₅₀=(10±0.5)μm

 

Collection Efficiency

σg=1.5±0.1